High Voltage
Test Unit HV 100-30 and HV 60-50
|
The COMET high voltage tester generates an AC
voltage of maximum 100/60 kV peak. The frequency is that of the line
voltage, either 50 or 60 Hz.
Depending on the selected position of the selector
switch, the voltage applied to the test object can be AC or in
the DC mode DC+ or DC- (half wave rectification).
The voltage applied to the item to be tested is measured in peak
units independent of the position selected at the selector switch.
In order to avoid damage to the capacitors to be tested, the DC
output voltage is limited to 50 kV. |
|
 |
The current ranges show a RMS value in the range of 30
and 1 mA for the current flowing through the device being tested. In the
AC mode this is true RMS value. In the case of DC this is a pulsating
direct current if a resistor is being tested and a periodic charge current
if a capacitor is in the test position.
In the 100 µA range the leakage current of a capacitor
can be measured. It must be noted that the DC voltage applied is
pulsed (half wave rectification). Therefore the indicated current is a
result of two currents, the leakage current and the charge current.
Depending on the capacitance value and the level of the leakage current,
the effect of the latter can be ignored.
Depending on the charge current and the leakage current
present, the voltage on the capacitor is more or less affected by a
ripple. The current shown is the average leakage current, depending on the
ripple experienced.
|
Specifications: |
Type HV 100-30 with LVO |
Type HV 60-50 with LVO |
|
With Low Voltage Option |
Part No. 91110 |
Part No. 91210 |
|
Electrical: |
|
|
|
Output Voltage
AC
DC |
0
- 100 kV peak 0 - 50 kV peak (incl. ripple) |
0
- 60 kV peak 0 - 50 kV peak (incl. ripple) |
|
Output Current
AC
DC |
0
- 30 mA rms 0 - 1 mA |
0
- 50 mA rms 0 - 1 mA |
|
Measurement Accuracy:
Voltage:
Current: |
2
% 20 % |
2
% 20 % |
Line Input Requirements
(to be specified at time of order) |
230 V / 115 V |
230 V / 115 V |
|
Power Consumption |
2150 VA |
2150 VA |
|
Frequency |
50/60 Hz |
50/60 Hz |
|
|
|
|
Modes of Operation: |
|
|
|
AC |
continuous/pulsed |
continuous/pulsed |
|
DC |
+/- polarity |
+/- polarity |
|
External (optional) |
-
voltage measurement - voltage source |
-
voltage measurement - voltage source |
Safety
Provisions:
- All high voltage components and
connections contained within one envelope
- Safety interlocks on all access
openings
- No high voltage connection outside the
safety enclosure
- HV zero start feature
- X-ray radiation shielding
- Overload protection
- Grounding bar
- Provisions for external
interlocks
|
|
LVO = Low
Voltage Option:
- Reduction of the output voltage to
approx. 20 %
- Increase of the sensitivity of the arc
detector
Optional Accessories:
- External measurement set, including:
HV cable 100 kV, 3 m (Part Number: 91500)
- Remote display unit,
including:
19" Einschub (Part Number: 91541)
Voltmeter (Part Number: 91542)
Amperemeter (Part Number: 91543) |
|